3 Using the Module
Installing Valve Heads
1 Switch off the module. 2 When unscrewing the transportation lock, push it back
until the last screw is removed - the valve rail is
spring-loaded.
3 Press on the valve dummy (1.) to release it (2.)
(spring-loaded valve rail).
4 Unscrew the valve dummy.
5 Remove the valve dummy from the valve drive.