1260 Infinity II Binary LC System - Manual and Quick Guide 37
Configuration Settings
2
Instrument Configuration
Instrument Configuration
Example shows an instrument configuration with a Diode Array Detector.
1 Set the switches of the Configuration switch at the rear of the module:
a All switches DOWN: module uses the default IP address 192.168.254.11.
b Switch 4 UP and others DOWN: module uses DHCP.
2 Enter the setup information (MAC / IP address and/or Instrument Name).
a Agilent OpenLab ChemStation (Configure Instrument):