EasyManuals Logo

Agilent Technologies G7129A Quick Guide

Agilent Technologies G7129A
82 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #37 background imageLoading...
Page #37 background image
1260 Infinity II Binary LC System - Manual and Quick Guide 37
Configuration Settings
2
Instrument Configuration
Instrument Configuration
Example shows an instrument configuration with a Diode Array Detector.
1 Set the switches of the Configuration switch at the rear of the module:
a All switches DOWN: module uses the default IP address 192.168.254.11.
b Switch 4 UP and others DOWN: module uses DHCP.
2 Enter the setup information (MAC / IP address and/or Instrument Name).
a Agilent OpenLab ChemStation (Configure Instrument):

Other manuals for Agilent Technologies G7129A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7129A and is the answer not in the manual?

Agilent Technologies G7129A Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7129A
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals