54 1260 Infinity II Binary LC System - Manual and Quick Guide
3
Quick Start Guide
Check Out the System
Check Out the System
Checkout Method
This is an examplary method for the Agilent InfinityLab LC Series.
Examplary configuration:
• Binary Pump G7112B
• Multisampler G7167A
• Multicolumn Thermostat G7116A
• Diode Array Detector WR G7115A
The RRLC checkout sample (5188-6529) serves as standard for systems with
600 bar and contains 100 ng/μL each of nine components dissolved in water /
acetonitrile (65/35). The nine components are:
• Acetanilide
• Acetophenone
• Propiophenone
• Butyrophenone
• Benzophenone
• Valerophenone
• Hexanophenone
• Heptanophenone
• Octanophenone
The Agilent isocratic checkout sample (01080-68704) serves as standard for
systems with 400 bar and contains each of four components dissolved in
methanol.
• Dimethylphthalate
• Diethylphthalate
• Biphenyl
• o-Terphenyl