Memory Clearing, Sanitization and Removal Procedures
Instrument Sanitization Procedures
Instrument Sanitization Procedures
This section includes flowcharts that describe how to sanitize an instrument by physical removal and
replacement of the Disk Drive.
Application License Key Storage
Note that License keys for all Applications are stored in EEPROM on the A7 Midplane Assembly (as
described in Item 8 of Table 5-1 on page 14). Therefore, when replacing the Disk Drive, you do not need to
back up and restore the license keys.
Replacement of Disk Drive
Refer to the flowchart in Figure 6-1 below for details of how to perform this procedure.
For details of how to archive or restore the instrument’s calibration files (Steps 3, 12 and 16 in the flowchart),
see Archiving and Restoring Factory Calibration Data Files” on page 25.
For details of how to remove the Disk Drive (Step 6), see “Disk Drive Removal Procedure” on page 27.
IMPORTANT When installing a replacement Disk Drive, ensure that the instrument software
revision on the replacement drive matches that of the original drive.