Installation Note E4440-90318 17
Installation Procedure
Install the Option Designator and License Key Number
NOTE The option designator 107 and the license key number must be entered
into instrument memory in addition to the correct firmware before the
hardware will function.
1. Locate the Option Upgrade Entitlement Certificate in the kit and follow the
directions to redeem it. You will receive an email that has the license key number
required for step 4.
2. Plug in instrument and power up. There may be alignment error messages since new
hardware was installed.
3. On the instrument front panel press:
System, More, until the Licensing softkey is
visible. Press
Licensing and Option. This will activate the alpha editor menu. Use the
front panel numerical keypad to enter the option designator 107. Press the
Enter key.
Note that 107 now appears on the
Option key.
4. Press
License Key. The license key number is a hexadecimal number that will require
the entry of both letters and numbers. The license key number appears on the license
key number certificate that you received when you redeemed your Option Upgrade
Entitlement Certificate. Use the alpha editor and the front panel numerical keypad to
enter the license key number. Your entry will appear in the active function area of the
display. If you make a typing error, use the backspace key to correct the error. Verify
the license key number you entered. Press
Enter, Activate License.
5. Cycle instrument power and allow instrument to perform the auto align routine.