Installation Note E4440-90318 19
Installation Procedure
Adjustments and Performance Tests Required
Adjustments Required
Performance Tests Required
The performance tests listed below are the minimum set that will verify the hardware
retrofit kit just installed is functioning correctly. Performing only these tests does not
guarantee the instrument meets all specifications.
NOTE A full calibration is required to assure the instrument meets all
specifications
Performance testing requires the use of the calibration software. The latest software
information and downloads are available at:
http://www.agilent.com/find/calibrationsoftware
End of installation
For assistance, get in touch with your nearest Agilent Technologies Sales and Service
Office. To find your local Agilent office access the following URL, or call the following
telephone number:
http://www.agilent.com/find/assist
1-800-452-4844 (8am-8pm EST)
Adjustments
Audio Step Gain and Calibrator adjustment
Audio Flatness adjustment
Performance Tests
Absolute Amplitude Accuracy
Resolution BW Switching
Display Scale Fidelity
Audio Distortion
Audio Amplitude Accuracy