XGS-600 Gauge Controller
32
DRAFT 4/
15/14
Figure 3-6 Sensor Setup Screen: Analog Card
See Table 3-1 on page 33 for default values. Fields that are unavailable have
- - -.
Sensor ID Use this field to select the sensor to view its setup parameters.
Displays the unique system-assigned gauge identification. Scroll
through this list to select a sensor and the remainder of fields on the
screen populate with its present settings.
Possible sensor ID’s are:
❑ HFIGx for Hot Cathode Ion Gauge boards using BA or UHV
sensors.
❑ CNVx for Convection boards (TC/ConvecTorr)
❑ IMGx for IMG Gauge boards
❑ AUXx for Analog Input Boards (Active Gauges)
where x indicates the number of the transducer attached. The num-
ber is system-assigned based on the sequence of board installation,
left to right (when viewed from the front). Convection boards have
two sensors per board. When more than 4 convection boards are
installed, the 10, 11, and 12 channels are designated CNVA,
CNVB, and CNVC, respectively.
User Label Use this field to enter your label for this sensor that appears on all
other screens and serial communication. This field has a maximum
of five (5) characters and any alphanumeric character is available
for use.
User Labels that begin with HFIG, CNV, IMG or AUX are not
allowed to avoid confusion with Sensor ID’s.
If no User Label is selected, the default is the Sensor ID. When a
System Reset is initiated, this label is reset to the sensor ID.