XGS-600 Gauge Controller
56
DRAFT 4/
15/14
Gauge Connector
Collector: SMB coax
Fil and Grid:AMP 5 pos Univ Mate-N-Lok
Remote I/O
Connector: 6 pin IEEE 1394 (Firewire type)
Inputs for Emission on/off and DEGAS on/off are user select-
able to be either Active LO or Active HI. Driving the input to
the active state turns on that function.
HI = +5 V max, 2.4 V min
LO = 0.8 V max, 0 V min
Filament Select input: Pin 1 open or driven HI selects Fil1.
Pin 1 grounded selects Fil2.
Status Output goes HI during error condition
Analog Output
Allowable Cable Length
up to 50’ using Agilent standard cable
up to 300’ using under 10’ Agilent standard cable + Agilent
extension cable
Table A-2 HFIG Gauge Specifications