XGS-600 Gauge Controller
DRAFT 4/15/14
71
Read Auto Fil Advance
State
#aa37 >xx
where:
xx = 00 is OFF
xx = 01 is ON
Set degas OFF #aa40cn >
Set degas ON #aa41cn >
Read degas status #aa42cn >xx
where:
xx = 00 is OFF
xx = 01 is ON
Read ion gauge gas
correction
#aa50cn >x.xxx
Set ion gauge gas
correction
#aa51cnx.xxx >
Read Emission current #aa52cn >x.xxx Least significant two
digits always zero.
Set Emission current #aa53cnx.xxx > Least significant two
digits will be ignored.
Read Sensitivity #aa54cn >xx.xx Always reads in Torr.
Least significant digit is
always zero.
Set Sensitivity #aa55cnxx.xx > This sensitivity value
must always be in Torr
due to xx.xx format
restriction. Note
however the value
displayed on front
panel is in Torr,
mbar, or Pascal.
Least significant digit
will be ignored.
Note: Will overwrite
any factory setting.
Table B-1 Serial Command Set (Continued)
Description Command
Response from
XGS-600 Notes