Column operating parameters: pH and temperature
Note: All silica-based packings have some solubility in pH >6 aqueous
mobile phases. When using silica-based columns at pH >6, best column
lifetime is obtained at lower temperatures (40 °C max) using low buffer
concentrations in the range of 0.01 to 0.02 M. Operating at extreme ends
of pH and temperature ranges will have a signicant impact on column
lifetime.
Phase
Recommended
pH Range
Maximum Operating
Temperature
Poroshell HPH pH 3.0 to 11.0 60 °C
Poroshell 120 SB-C18,
StableBond C18,
ZORBAXRx-C8,
ZORBAX SB-Aq,
ZORBAX SB-Phenyl,
SB-C3, Poroshell 300,
300SB-C3
pH 1.0 to 8.0 90 °C (StableBond C18
and Poroshell 120 SB-C18)
80 °C (Rx-C8 and SB-Aq,
SB-C3 and SB-Phenyl)
70 °C for pH <5.0; 40°C
for pH 5.0 to 8.0 (Poroshell
300 and 300SB-C3)
ZORBAX Eclipse Plus C18
and C8, Eclipse XDB-C8,
XDB-C18, XDB-Phenyl,
Poroshell 120 EC-C18 and
EC-C8, ZORBAX Bonus-RP,
HC-C18(2) and TC-C18(2)
pH 2.0 to 9.0 60 °C
ZORBAX Extend C18 pH 2.0 to 11.5 60 °C (40 °C at high pH)
Poroshell 120 PFP,
Poroshell 120 Phenyl-Hexyl,
ZORBAX ODS,
ZORBAXPhenyl,
ZORBAXEclipse Plus
Phenyl-Hexyl,
ZORBAX XDB-CN,
ZORBAXEclipse PAH,
AllPursuit and Polaris
bonded phases
pH 2.0 to 8.0 60 °C
ZORBAX Eclipse PAH pH 2.0 to 8.0 60 °C at pH <6.0
ZORBAX TMS pH 2.0 to 7.0 60 °C

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies ZORBAX Eclipse Plus C18 and is the answer not in the manual?

Agilent Technologies ZORBAX Eclipse Plus C18 Specifications

General IconGeneral
BrandAgilent Technologies
ModelZORBAX Eclipse Plus C18
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals