7
Triggered Burst
Each active edge of the trigger signal will produce one burst of the waveform.
Carrier Waveforms:
Sine, Square, Ramp, Arb, Pulse, PRBS
Maximum Carrier Frequency: 10MHz (finite cycles), 50MHz [25MHz] (infinite), subject to carrier
waveform.
Number of Cycles:
1 to 16,777215 and infinite.
Trigger Repetition Rate: 2mHz to 1MHz internal
dc to 1MHz external.
Trigger Signal Source: Internal from keyboard, trigger generator, or other channel.
External from TRIG IN or remote interface.
Trigger Start/Stop Phase: -360.0 to +360.0 degrees, 0.1 degree resolution, subject to carrier
waveform
Gated
Waveform will run while the Gate signal is true and stop while false.
Carrier Waveforms:
Sine, Square, Ramp, Arb, Pulse, Noise, PRBS
Maximum Carrier Frequency:
10 MHz, subject to carrier waveform
Trigger Repetition Rate: 2mHz to 1MHz internal
dc to 1MHz external.
Gate Signal Source: Internal from keyboard, trigger generator, or other channel.
External from TRIG IN or remote interface.
Gate Start/Stop Phase: -360.0 to +360.0 degrees, 0.1 degree resolution, subject to carrier
waveform
Sweep
Frequency sweep capability is provided for both standard and arbitrary waveforms.
Carrier Waveforms:
All standard and arbitrary except pulse.
Sweep Mode:
Linear or logarithmic, triggered or continuous.
Sweep Direction:
Up, down, up/down or down/up.
Sweep Range: From 1µHz to 50MHz [25MHz], subject to carrier waveform. Phase
continuous. Independent setting of the start and stop frequency.
Sweep Time:
1ms to 500s (6 digit resolution).
Marker:
Variable during sweep.
Sweep Trigger Source: The sweep may be free run or triggered from the following sources:
Internal from keyboard, trigger generator, or other channel.
Externally from TRIG IN input or remote interface.
Trigger Generator
Internal source 2mHz to 1MHz square wave adjustable in 1us steps, 9 digit resolution. Available for
external use from the SYNC OUT socket.