3.4 THERAPY MODE
SELECTION SCREEN
This is the THERAPY MODE SELECTION SCREEN.
To utilize Gradient Mode therapy, press UP or DOWN
to scroll to GRADIENT MODE. Press SELECT to conrm and
proceed to 3.5.
To utilize Pressure Mode, press UP or DOWN to scroll
to PRESSURE MODE. Press SELECT to conrm and proceed
to 3.8
To go back to the MAIN MENU SCREEN, press UP or
DOWN to scroll to BACK TO MAIN MENU. Press SELECT
to conrm.
3.5 GRADIENT MODE SET-UP
SCREEN
NOTE: Press UP or DOWN to scroll to MORE or
PREVIOUS to view additional settings options.
The gradient cannot be changed and is xed at 4%.
To set the Pressure, press UP or DOWN to scroll to
PRESSURE. Press SELECT to conrm. The pressure will
begin blinking. Press UP or DOWN to scroll from 30-80
mmHg, in increments of 1 mmHg. Press SELECT to conrm.
The pressure will be updated and stop blinking.
To set the Cycle Time, press UP or DOWN to scroll to
CYCLE TIME. Press SELECT to conrm. The cycle time will
begin blinking. Press UP or DOWN to scroll to 40 or
60 seconds. Press SELECT to conrm. The cycle time will be
updated and stop blinking.
GRADIENT MODE
PRESSURE MODE
BACK TO MAIN MENU
GRADIENT: 4%
PRESSURE: 45mmHg
CYCLE TIME: 60 sec
MORE
Operating Instructions (continued)
PREVIOUS
THERAPY TIME: 60 min
SKIP CHAMBER: OFF
MORE
PREVIOUS
LPT MODE: CLEARING
BACK TO MAIN MENU
BEGIN TREATMENT