13
A8IFU.0001 REV C
3.5 To set the Therapy Time, press UP or DOWN to scroll
to THERAPY TIME. Press SELECT to conrm. The therapy
time will begin blinking. Press UP or DOWN to scroll
to 15, 30, 45, or 60 minutes. Press SELECT to conrm. The
cycle time will be updated and stop blinking.
To select a chamber in which to skip pressure, press UP
or DOWN to scroll to SKIP CHAMBER. Press SELECT to
conrm. The skip chamber setting will begin blinking. Press
UP or DOWN to scroll through the settings, including
OFF, 1, 2, 3, 4, 5, 6, 7, or 8. Press SELECT to conrm. The
skip chamber setting will be updated and stop blinking.
To operate an optional Lymphedema Preparation Treatment
(LPT) mode before Gradient Mode therapy, press UP or
DOWN to scroll to LPT MODE. Press SELECT to conrm.
The LPT Mode option will begin blinking. Press UP or
DOWN to scroll through the settings, including OFF,
CLEARING, and FOCAL.
If you select FOCAL THERAPY, proceed to 3.6.
If you select CLEARING THERAPY, proceed to 3.7.
If you do not wish to run an LPT Mode option, scroll to OFF.
Press SELECT to conrm. The LPT Mode setting will be
updated and stop blinking.
To go back to the MAIN MENU SCREEN, press UP or
DOWN to scroll to BACK TO MAIN MENU. Press SELECT
to conrm.
To begin treatment therapy, press UP or DOWN to
scroll to BEGIN TREATMENT. NOTE: If this is the rst time
the device is being used, you may be asked to select the
number of sleeves to be used for therapy. Refer to 3.2 for
additional instructions.
If this is not the rst time the device is being used, you may
not be asked to select the number of sleeves. Treatment will
begin immediately.
Operating Instructions (continued)