15
A8IFU.0001 REV C
3.8 PRESSURE MODE SET-UP
SCREEN
INDIVIDUAL PRESSURE
SET-UP SCREEN
NOTE: Press UP or DOWN to scroll to MORE or
PREVIOUS to view additional settings options.
To set the Pressure in each chamber, press UP or
DOWN to scroll to SET PRESSURE. Press SELECT to
conrm. NOTE: Refer to the INDIVIDUAL PRESSURE SET-UP
SCREEN below.
To set the pressure in Chamber 1, press UP or
DOWN to scroll to CHAMBER 1. The pressure setting will
begin blinking. Press UP or DOWN to scroll from 35-80
mmHg, in increments of 1 mmHg. Press SELECT to conrm.
The percentage will be updated and stop blinking. Repeat
these steps to set the pressure in the rest of the chambers.
To save your changes, press UP or DOWN to scroll to
SAVE/EXIT. Press SELECT to conrm.
To set the Cycle Time, press UP or DOWN to scroll to
CYCLE TIME. Press SELECT to conrm. The cycle time will
begin blinking. Press UP or DOWN to scroll to 40 or
60 seconds. Press SELECT to conrm. The cycle time will be
updated and stop blinking.
To set the therapy Time, press UP or DOWN to scroll to
THERAPY TIME. Press SELECT to conrm. The therapy time
will begin blinking. Press UP or DOWN to scroll to 15, 30,
45, or 60 minutes. Press SELECT to conrm. The cycle time
will be updated and stop blinking.
To select a chamber in which to skip pressure, press UP
or DOWN to scroll to SKIP CHAMBER. Press SELECT to
conrm. The skip chamber setting will begin blinking. Press
UP or DOWN to scroll through the settings, including
OFF, 1, 2, 3, 4, 5, 6, 7, or 8. Press SELECT to conrm. The skip
chamber setting will be updated and stop blinking.
To go back to the MAIN MENU SCREEN, press UP or
DOWN to scroll to BACK TO MAIN MENU. Press SELECT
to conrm.
To operate an optional Lymphedema Preparation Treatment
(LPT) mode before Pressure Mode therapy, press UP or
DOWN to scroll to LPT Mode. Press SELECT to conrm.
The LPT Mode option will begin blinking. Press UP or
DOWN to scroll through the settings, including OFF,
CLEARING, and FOCAL.
PREVIOUS
LPT MODE: CLEARING
BEGIN TREATMENT
PREVIOUS
SKIP CHAMBER: OFF
BACK TO MAIN MENU
MORE
SET PRESSURE: 1-8
CYCLE TIME: 60 sec
THERAPY TIME: 60 min
MORE
CHAMBER 1: 60mmHg
CHAMBER 2: 57mmHg
CHAMBER 3: 54mmHg
MORE
PREVIOUS
CHAMBER 4: 51mmHg
CHAMBER 5: 48mmHg
MORE
PREVIOUS
CHAMBER 8: 39mmHg
SAVE/EXIT
PREVIOUS
CHAMBER 6: 45mmHg
CHAMBER 7: 42mmHg
MORE
Operating Instructions (continued)