AS-A—AS-P Series Instruction Manual A-3
AS-A
Oxygen output: 20-25 SCF/hr*
Oxygen Delivery Pressure: 50 psig (max)
345 kPa (max)
Oxygen purity: 90% (min)
Oxygen dew point: -73°C (-100°F)
Dimensions: 11 x 13 x 57 in. (L x W x H)
28 x 33 x 145 cm (L x W x H)
Approximate weight: 140 lb
64 kg
Feed Air Requirements
Pressure: 90 psig (620 kPa) (min)
Temperature: 50°C (122°F) (max)
Feed Air Regulator Setting***: 70-75 psig
Power requirements**: 120 (± 10%) VAC, 50/60 Hz, Single Phase,
3 Amp
240 10%) VAC, 50/60 Hz, Single Phase,
1 Amp
Pressure Switch Setting***: Minimum: 46 - 48 psig
Maximum: 54 - 56 psig
Other Specifications:
Pollution Degree 2
Installation Category II
Altitude 2000 m
*SCF (standard cubic foot) gas measured at 1 atmosphere and 70°F.
**Check the label on the control panel for appropriate power supply.
***Data may vary as per the customer’s requirements.