AirSep Corporation
4.0 Maintenance
4.1 GENERAL
The Compressor Inlet Filter should be changed approximately every 4,000 hours of
operation. Filter change frequency is dependent on environmental conditions and
may vary.
Compressors require a rebuild after 5,000 to 12,000 hours of operation depending
on environmental conditions.
See the Service Parts section for information on obtaining lters and compressor
rebuild kits.