4-7. Releasing an exposure
1) Pick up the exposure release switch.
NOTE
Ensure radiation protection.
Maintain maximum possible distance from the source.
☞ Refer to page 8, 2-1. Radiation protection
2)
Press the exposure release switch to
the prep position ( i.e. first position )
and hold for 1∼2 seconds,
then READY indicator will glows on the
operating control panel.
Do not press the exposure release
switch to the exposure position.
( i.e. second position )
NOTE
After the X-ray tube has been warmed up (approx. 1∼2sec.), the exposure
can be released without delay. This helps to eliminate motion artifacts.
-------------------------------------------------------------
AJEX240H / Operating instructions -23-
AXR-240H-MN01 (E) / Rev. 03