What does error D08 no connect mean on my Alcatel Vacuum Technology Adixen ATP?
- MMrs. Bridget LeeSep 9, 2025
If the Alcatel Vacuum Technology Water Pump displays 'D08 : NO CONNECT', check the cable connection to the pump.
What does error D08 no connect mean on my Alcatel Vacuum Technology Adixen ATP?
If the Alcatel Vacuum Technology Water Pump displays 'D08 : NO CONNECT', check the cable connection to the pump.
What does error D26 no connect mean on my Alcatel Vacuum Technology Adixen ATP?
If the Alcatel Vacuum Technology Water Pump displays 'D26 : NO CONNECT', check the cable controller electrical connection.
What does error D02 DLY software mean on my Alcatel Vacuum Technology Adixen ATP Water Pump?
If the Alcatel Vacuum Technology Water Pump displays 'D02 : DLY SOFTWARE' due to a soft counter default, reinitialize the controller with the main switch (0/1).
What does error D01 EXT safety mean on my Alcatel Vacuum Technology Adixen ATP Water Pump?
If the Alcatel Vacuum Technology Water Pump displays 'D01 : EXT SAFETY' because the external security contact on the REMOTE CONTROL connector is activated, test the external safety devices (contact 31 - 32); repair the fault and press START to restart.
Brand | Alcatel Vacuum Technology |
---|---|
Model | Adixen ATP Series |
Category | Water Pump |
Language | English |
Explains how turbomolecular pumps work and their installation within a system.
Provides detailed technical specifications for standard, corrosive, and HPC ATP pumps.
Illustrates dimensional drawings and tables for ATP 80, 100, 150, 400, and 900 pumps.
Details electrical and dimensional characteristics for ACT 200T, 600T, and 1000T controllers.
Provides essential safety precautions for handling, installation, and operation.
Details electrical connections for ACT 200 T and 600 T controllers.
Explains wiring for remote control inputs using voltage or external contacts.
Guides through controller self-test, initialization, and parameter setting.
Details electrical connections for the ACT 1000 T controller.
Provides detailed explanations of RS commands for pump control and configuration.
Covers initial pump start-up procedure and critical safety instructions.
Lists incidents, causes, consequences, and remedies for ACT 200T/600T.
Lists incidents, causes, consequences, and remedies for ACT 1000 T.