EasyManuals Logo

Alere Cholestech LDX User Manual

Alere Cholestech LDX
50 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #14 background imageLoading...
Page #14 background image
12 Alere Cholestech LDX
®
System User Manual
Setting the Conguration Menu
Before using your Alere Cholestech LDX
®
Analyzer, you should
enter the Conguration Menu and set it to best meet your needs.
The following items can be customized:
Language English, Italian, Spanish, German, French
Units mg/dL=milligrams per deciliter
mmol/L=millimoles per liter
NOTE: U/L (units per liter) and mg/L (milligrams
per liter) will be reported when the Conguration
Menu is set for either UNITS=mg/dL or
UNITS=mmol/L.
Sample Whole B.=Venous or capillary whole blood
Serum=Serum or plasma, calibration verication
or prociency testing material
NOTE: Check the package insert for the test
cassette you are using to determine the correct
SAMPLE mode when testing quality control or
calibration verication material.
LDL/HDL ratio No = LDL/HDL function turned off, TC/HDL
function turned on
Yes = LDL/HDL function turned on, TC/HDL
function turned off
Printer On=printer function turned on
Off=printer function turned off
Number of
Copies
1=1 printed label
2=2 printed labels
3=3 printed labels
Risk CHD Risk=Framingham Risk on
Off=Risk assessment option is turned off
Date Day of week, month, day, and year=English
Day of week, day, month, and year=Alternate
language settings
Time Hour, minutes, seconds, AM/PM=English
Hour, minutes, seconds, (24 hour)=Alternate
language settings

Other manuals for Alere Cholestech LDX

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Alere Cholestech LDX and is the answer not in the manual?

Alere Cholestech LDX Specifications

General IconGeneral
BrandAlere
ModelCholestech LDX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals