16 Alere Cholestech LDX
®
System User Manual
Alere Cholestech LDX
®
Optics Check
An Alere Cholestech LDX
®
Optics Check Cassette is supplied with
the analyzer. It should be used to check the optical system of the
analyzer. Store the Alere Cholestech LDX
®
Optics Check Cassette
at room temperature in the case provided. Do not touch the
reaction bar or allow it to become wet, dirty or scratched. Do not
use a damaged or expired Alere Cholestech LDX
®
Optics Check
Cassette.
Run the Alere Cholestech LDX
®
Optics Check Cassette:
Once each day before patient samples are tested.
After the Alere Cholestech LDX
®
System has been moved
or serviced.