6. Squeeze the nger gently to obtain a large drop of blood.
Wipe away this rst drop of blood, as it may contain
tissue uid.
7. Squeeze the nger gently again while holding it
downward until a second large drop of blood forms.
Do not milk the finger. The puncture should provide a
free-owing drop of blood.
8. Hold the capillary tube horizontally or at a slightly
descending angle by the end with the plunger. Touch it to
the drop of blood without touching the skin. The tube will
ll by capillary action up to the black mark. Do not collect
air bubbles. If it is necessary to collect another drop of
blood, wipe the nger with gauze then massage again from
base to tip until a large drop of blood forms.
9. Fill the capillary tube within 10 seconds.
10. Wipe off any excess blood from the nger and have the
patient apply pressure to the puncture until the bleeding