AliveCor, Inc. For support, please email support@alivecor.com or visit www.alivecor.com 32
Recommended separation distances between
portable and mobile RF communications equipment and Kardia Mobile
Kardia Mobile is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of Kardia Mobile can help prevent electromagnetic interference by maintaining a minimum distance
between portable and mobile RF communications equipment (transmitters) and Kardia Mobile as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m)
can be determined using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
P
V
d ]
5.3
[
1
P
E
d ]
5.3
[
1
P
E
d ]
7
[
1

Related product manuals