AIR VELOCITY METER
ALNOR
®
MODEL
AVM430/AVM430A
AIRFLOW™ INSTRUMENTS
MODEL TA430/TA430A
OPERATION AND SERVICE MANUAL
P/N 1980583, REV F
SEPTEMBER 2014
w ww. . com
information@itm.com1.800.561.8187

Questions and Answers

  • J
    judypearsonAug 10, 2025
    What to do if Alnor AVM430 Measuring Instruments have no display?
    • J
      jenniferwilsonAug 11, 2025
      If your Alnor Measuring Instruments display is blank, the unit might not be turned on. Ensure the device is switched on. Alternatively, the batteries could be low or dead; try replacing them or plugging in the AC adapter. Another possibility is dirty battery contacts, so try cleaning them.
  • W
    William WilliamsAug 18, 2025
    Why does Alnor AVM430 Measuring Instruments show an error message?
    • D
      Daniel RichardsonAug 18, 2025
      If your Alnor Measuring Instruments show an error message, it could mean that the memory is full. If you need the data, download it first, and then delete all the memory. Alternatively, there may be a fault in the instrument, and factory service is required.
  • J
    Jessica HamiltonAug 25, 2025
    How to fix unstable velocity reading on Alnor Measuring Instruments?
    • D
      David BoyleAug 26, 2025
      If the velocity reading on your Alnor Measuring Instruments is unstable, it might be due to fluctuating flow. Try repositioning the probe in a less turbulent flow or use a longer time constant.
  • S
    sierranicholsSep 2, 2025
    What to do if Alnor Measuring Instruments keypad is not responding?
    • K
      Kimberly HarrisSep 2, 2025
      If your Alnor Measuring Instruments keypad is not responding, it may be locked. Unlock the keypad by pressing the keys simultaneously.

Summary

Setting-up

Using the Telescoping Probe

Using the probe, ensuring sensor exposure and orientation for accurate readings.

Connecting to a Computer

How to connect the instrument to a computer using a USB cable for data download.

Operation

Keypad Functions

Explanation of the functions of the device's keypad buttons.

Menus

Overview of the device's menu structure and navigation.

Display Setup

How to configure parameters displayed on the running screen.

Data Logging

Managing data logging features, including measurements, deletion, and memory.

LogDat2 Downloading Software

Instructions for installing and using the software to download data.

Maintenance

Recalibration

Information on annual recalibration and field adjustments for maintaining accuracy.

Troubleshooting

Troubleshooting the Alnor AVM430;Airflow Instruments TA430

Table listing common symptoms, causes, and corrective actions for device issues.

Specifications

Velocity Specifications

Specifications for velocity measurement range, accuracy, and resolution.

Temperature Specifications

Specifications for temperature measurement range, accuracy, and resolution.

Alnor AVM430 Specifications

General IconGeneral
DisplayLCD
Air Velocity Range0.2 to 30 m/s
Temperature Range-10 to 60 °C
Temperature Accuracy±0.3 °C
Resolution0.01 m/s, 0.1 °C
Operating Temperature-10 to 60 °C
Storage Temperature-20 to 70 °C
Power4 AA batteries
Humidity Range0 to 100% RH
Accuracy±2% of reading ±0.1 m/s