31
℃
℃
pi,pi,pi,pi,pi,
pi,pi,pi,pi,pi,
During exhaust
PATTERN
MON DAY
HR MIN
EXHAUST
START
TIMER
STERILIZE
START
STOP
COMPLETE
O
.
T
.
TEMP
TIME
ENTER
HEAT
EXHAUST
PATTERN
MON DAY
HR MIN
EXHAUST
START
TIMER
STERILIZE
START
STOP
O
.
T
.
TEMP
TIME
ENTER
HEAT
EXHAUST
④
⑤
1
0
7
0
0
0
8
0
0
0
0
Lighting : Lid lock can not be released yet.
Press STOP KEY.
Extinction : Lid lock can be released.
⑥
5.9 Operation complete
④ After sterilization time has elapsed, complete alarm sounds then power to heater is cut off. In case
of setting automatic exhaust ON as above chart, automatic exhaust starts after temperature in
the chamber reaches to automatic exhaust starting temperature to decrease pressure forcibly in
chamber.
(In case of automatic exhaust OFF, EXHAUST LAMP does not blink and autoclave does not exhaust
steam automatically. When you press EXHAUST KEY, autoclave exhausts steam.)
COOLING-1 PROCESS LAMP blinks from sterilization temperature to 100℃.
After that COOLING-2 PROCESS LAMP brinks from 99℃ to completion.
⑤ When TEMPERATURE DISPLAY indicates 80℃, complete alarm sounds and COMPLETE LAMP
lights up.
⑥ Press STOP KEY to release the lid lock. Then the lid can be opened and COMPLETE
LAMP turns off.
※In case of interruption of operation, the lid can not be opened till TEMPERATURE
DISPLAY indicates below 80℃ too.