47
Do not scratch heater in bottom of chamber when cleaning.
The water which infiltrates into heater may cause breakdown.
When cleaning, check the chamber temperature if it is cold enough, or it may cause burn injury.
Do not pull out the lid gasket forcibly.
It may cause steam leakage.
Use base plate hook when unloading base plate.(CL-40L only)
④ Check of power supply and its peripherals
● Power connector
1.Make sure if the power cable does not have any damage and there are not
heavy things on the cable.
2.Make sure if the power connector of the autoclave is well connected to the
power receptacle.
3.If the dust is attached with power connector, clean up them.
● Breaker
1.Check if the breaker(power switch) is shut off
by pushing the test button by a sharp object (pen etc.).
Do not use autoclave if power cable and/or power connector are damaged.
Contact local distributor, or it may cause electric shock.
If the dust is attached with power connector, clean up them.
It may cause fire by heat.
Do not touch the power connector and breaker by wet hand.
It may cause electric shock.
CAUTION
WARNING