User Manual i4MC&V/EN M/C
iSTAT i4MC&V
Page 10
1.3.3 Software:
The
same software is used for configuring the device as on all iSTAT communicating
products.
QDSP Standard for setting and browser software
QDSP also offers additional features such as upgrading from a secure web site for
both the QDSP software and the transducer firmware.
1.4 Measurements
i4MC/V is id
eally suited to applications where continuous monitoring of a single phase
system is required.
1.5 Hardware features
i4MC/V has a
Red LED indicator showing that power is on.
i4MC/V has a universal auxiliary supply and an auto ranging current or voltage measurement
input so that it can be used in most site conditions without the need to specify this
information at the order stage.
1.6 Communications
i4MC/V
can be supplied with either RS232 or RS485 communications via screw terminals,
which has to be defined at the time of ordering. The communications are used for
programming the transducer or for monitoring measurements.
In addition a USB port (mini type) is always provided on the underside of the i4MC/V for
programming the transducer before installation. The USB port cannot be used when the
i4MC/V is installed on the DIN Rail.
Note: The USB port has only Basic Insulation and can only be used
when there is no wiring connected to the main terminals.
Note: The Cover over the USB connector must be fitted prior to
installation or storage, if not the Warranty on the product will be void.
1.7 Inputs and Outputs
i4MC/V is fitted with one analogue output circuit which is fully programmable over the ranges
0 - 20mA and 0 - 10V.