Alarm Messages
The unit is in constant self-diagnosis and shows a variety of
messages on the four-line display to warn of error conditions:
The machine has detected that the CR2 relay has failed. You will
need to return the unit to AMD for repair. Please call (303) 755-0112
for further instructions.
The machine has detected an unusually high internal temperature.
Ensure there is nothing blocking the fan or that the unit is not in
direct sunlight. Turn o the unit and wait 30 minutes to turn back on.
The machine has detected one of the following: the power pack
has failed, or the wrong power pack is plugged into the machine,
or the voltage regulator in the external power pack has failed, or
the normal 20-volt supply voltage has exceeded 22.1 volts for some
reason.
There is too much salt in the water. Empty the foot tub and add
fresh water. Add salt slowly to the water. Typically, ¼ teaspoon is the
most that is required. Be sure that the voltage does not drop below
11.