102 23-03-2009 105446 06
THERMAL SPECIFICATIONS ATC-156 A/B
Influence from load with
Ext. Reference :
0.01°C (0.02°F) @155°C (311°F)
0.01°C (0.02°F) @-24°C (-11.2°F)
Long term drift (1 year) :
±0.05°C (±0.09°F)
Calibration accuracy (test limit) ±0.05°C (0.09°F)
Temperature coefficient 5ppm/°C (0-40°C) / (32-104°F)
Stability ±0.01°C ( ±0.02F)
Reference accuracy ±0.02°C (±0.04°F)
Total accuracy ±0.10°C (±0.18°F)
Heating time incl. insert -24°C / -11.2°F to 23°C / 73.4°F : 4 min.
23°C / 73.4°F to 100°C / 212°F: 9 min.
23°C / 73.4°F to 155°C / 311°F : 19 min.
100°C / 212°F to 155°C / 311°F: 10 min.
-24°C / -11.2°F to 155°C / 311°F : 23 min.
Time to stability 7 min.
Cooling time incl. insert 155°C / 311°F to 100°C / 212°F: 4 min.
155°C / 311°F to 23°C / 73.4°F : 13 min.
100°C / 212°F to 23°C / 73.4°F: 9 min.
23°C / 73.4°F to 0°C / 32°F: 6 min.
23°C / 73.4°F to –24°C / -11.2°F : 29 min.
0°C / 32°F to –20°C / -4°F: 13 min.
155°C / 311°F to –24°C / -11.2°F : 42 min.

Table of Contents

Other manuals for Ametek JOFRA ATC-125 A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek JOFRA ATC-125 A and is the answer not in the manual?

Ametek JOFRA ATC-125 A Specifications

General IconGeneral
BrandAmetek
ModelJOFRA ATC-125 A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals