105446 06 23-03-2009 109
THERMAL SPECIFICATIONS –
LIQUID BATHS
ATC-250 A/B
Well specifications 40 mm / 1.57 inch axial homogeneity :
0.04°C / 0.07°F @250°C / 482°F
0.04°C / 0.07°F @140°C / 284°F
0.03°C / 0.05°F @50°C / 122°F
50 mm / 1.97 inch axial homogeneity :
0.04°C / 0.07°F @250°C / 482°F
0.04°C / 0.07°F @140°C / 284°F
0.04°C /0.07°F @50°C / 122°F
60 mm / 2.36 inch axial homogeneity :
0.04°C / 0.07°F @250°C / 482°F
0.04°C / 0.07°F @140°C / 284°F
0.04°C /0.07°F @50°C / 122°F
70 mm / 2.76 inch axial homogeneity :
0.05°C / 0.09°F @250°C / 482°F
0.05°C / 0.09°F @140°C / 284°F
0.04°C / 0.07°F @50°C / 122°F
80 mm / 3.15 inch axial homogeneity :
0.10°C / 0.18°F @250°C / 482°F
0.05°C / 0.09°F @140°C / 284°F
0.04°C / 0.07°F @50°C / 122°F
Difference between borings :
0.025°C / 0.045°F
Influence from load :
0.10°C / 0.18°F @250°C / 482°F
Influence from load with
Ext. Reference :
0.02°C / 0.04°F @250°C / 482°F
Difference between insert / oil :
0.40°C / 0.18°F

Table of Contents

Other manuals for Ametek JOFRA ATC-125 A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek JOFRA ATC-125 A and is the answer not in the manual?

Ametek JOFRA ATC-125 A Specifications

General IconGeneral
BrandAmetek
ModelJOFRA ATC-125 A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals