108 23-03-2009 105446 06
THERMAL SPECIFICATIONS –
DRY - BLOCK
ATC-250 A/B
Calibration accuracy (test limit) ±0.13°C / ±0.23°F
Temperature coefficient 5ppm/°C (0-40°C) (32-104°F)
Stability ±0.02°C / ±0.04°F
Reference accuracy ±0.06°C (±0.11°F)
Total accuracy ±0.28°C (±0.50°F)
Heating time incl. insert 28°C / 82°F to 100°C / 212°F : 3 min.
50°C / 122°F to 100°C / 212°F : 2 min.
100°C / 212°F to 250°C / 482°F : 9 min.
Time to stability 15 min.
Cooling time incl. insert 250°C / 482°F to 100°C / 212°F : 27 min.
100°C / 212°F to 50°C / 122°F : 27 min.
100°C / 212°F to 28°C / 82°F : 55 min.
THERMAL SPECIFICATIONS –
LIQUID BATHS
ATC-250 A/B
Maximum temperature 250°C / 482°F
Minimum temperature 10°C / 50°F @ ambient temperature 0°C / 32°F
33°C / 91°F @ ambient temperature 23°C / 73.4°F
50°C / 122°F@ ambient temperature 40°C /
104°F