118 23-03-2009 105446 06
INPUT SPECIFICATIONS
ACCURACY IN C/F
±0.48°C(±0.86°F) @ 1820°C(3308°F)
Accuracy thermocouple type N
(excluding sensor accuracy)
±0.20°C(±0.35°F) @ -90°C(-130°F)
±0.16°C(±0.29°F) @ -50°C(-58°F)
±0.15°C(±0.27°F) @ 0°C(32°F)
±0.14°C(±0.24°F) @ 155°C(311°F)
±0.14°C(±0.25°F) @ 320°C(608°F)
±0.16°C(±0.28°F) @ 650°C(1202°F)
±0.17°C(±0.31°F) @ 800°C(1472°F)
Accuracy thermocouple type XK
(excluding sensor accuracy)
±0.09°C(±0.16°F) @ -90°C(-130°F)
±0.07°C(±0.13°F) @ -50°C(-58°F)
±0.06°C(±0.11°F) @ 0°C(32°F)
±0.06°C(±0.12°F) @ 155°C(311°F)
±0.07°C(±0.13°F) @ 320°C(608°F)
±0.11°C(±0.19°F) @ 650°C(1202°F)
±0.12°C(±0.22°F) @ 800°C(1472°F)
Accuracy thermocouple type U
(excluding sensor accuracy)
±0.16°C(±0.29°F) @ -90°C(-130°F)
±0.12°C(±0.21°F) @ -50°C(-58°F)
±0.10°C(±0.18°F) @ 0°C(32°F)
±0.09°C(±0.17°F) @ 155°C(311°F)
±0.09°C(±0.17°F) @ 320°C(608°F)
±0.10°C(±0.19°F) @ 600°C(1112°F)
Accuracy thermocouple type L
(excluding sensor accuracy)
±0.08°C(±0.14°F) @ -50°C(-58°F)
±0.08°C(±0.14°F) @ 0°C(32°F)
±0.08°C(±0.14°F) @ 155°C(311°F)
±0.10°C(±0.18°F) @ 320°C(608°F)
±0.13°C(±0.23°F) @ 600°C(1112°F)
±0.14°C(±0.25°F) @ 900°C(1652°F)
Accuracy automatic cold junction
compensation
±0.40°C (±0.72°F) @ ambient temperature 0°C
to 40°C (32°F to 104°F).
Accuracy RTD Pt1000
(excluding sensor accuracy)
±0.043°C(±0.077°F) @ -90°C(-130°F)
±0.046°C(±0.083°F) @ -50°C(-58°F)