110 2014-07-30 127139 13
Remove the insulation plug while heating up.
It is very important that humidity in the well is removed to
prevent corrosion and frost expansion damages.
The stirring magnet and sensor basket (RTC-158/250 A/B/C
only) – When removed from the calibrator the items must be
cleaned thoroughly with dry disposable paper towels to avoid
spilling of fluid. Drops of silicone oil can be removed by using
special cleaning oil.
The sensor-under-test (RTC-158/250 A/B/C only) – When
removed from the calibrator the sensor must be cleaned
thoroughly with dry disposable paper towels to avoid spilling of
fluid. Drops of silicone oil can be removed by using special
cleaning oil.
Caution…
Avoid getting silicone oil on the clothes. It is impossible to
wash off.
7.2 Adjusting and calibrating the instrument
You are advised to return the calibrator to AMETEK Denmark A/S or
another accredited laboratory at least once a year for calibration.
Alternatively, you can calibrate/adjust the calibrator yourself using the
AmeTrim Adjust and Calibration Software.
Please refer to the Ametrim User Manual also found on the installation
USB memory stick.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Ametek JOFRA RTC-156 A and is the answer not in the manual?

Ametek JOFRA RTC-156 A Specifications

General IconGeneral
BrandAmetek
ModelJOFRA RTC-156 A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals