127139 13 2014-07-30 139
THERMAL SPECIFICATIONS RTC-159 A/B/C
±0.050°C/0.09°F @ 125°C/257°F to 155°C/311°F
50 mm / 1.97 inch axial homogeneity:
±0.100C/0.18°F @ -100°C/-148°F to 23°C/73.4°F
50 mm / 1.97 inch axial homogeneity:
±0.050°C/0.09°F @ 23°C/73.4°F to 125°C/257°F
50 mm / 1.97 inch axial homogeneity:
±0.050°C/0.09°F @ 125°C/257°F to 155°C/311°F
60 mm / 2.36 inch axial homogeneity:
±0.150C/0.27°F @ -100°C/-148°F to 23°C/73.4°F
60 mm / 2.36 inch axial homogeneity:
±0.050°C/0.09°F @ 23°C/73.4°F to 125°C/257°F
60 mm / 2.36 inch axial homogeneity:
±0.050°C/0.09°F @ 125°C/257°F to 155°C/311°F
70 mm / 2.76 inch axial homogeneity:
±0.200C/0.36°F @ -100°C/-148°F to 23°C/73.4°F
70 mm / 2.76 inch axial homogeneity:
±0.050°C/0.09°F @ 23°C/73.4°F to 125°C/257°F
70 mm / 2.76 inch axial homogeneity:
±0.050°C/0.09°F @ 125°C/257°F to 155°C/311°F
80 mm / 3.15 inch axial homogeneity:
±0.400C/0.72°F @ -100°C/-148°F to 23°C/73.4°F
80 mm / 3.15 inch axial homogeneity:
±0.050°C/0.09°F @ 23°C/73.4°F to 125°C/257°F
80 mm / 3.15 inch axial homogeneity:
±0.050°C/0.09°F @ 125°C/257°F to 155°C/311°F
Difference between borings :
0.01°C / 0.02°F
Influence from load :
0.20°C / 0.36°F @ -100°C / -148°F
0.05°C / 0.09°F @ 155°C / 311°F
Influence from load with Ext. Reference :
0.01°C / 0.02°F
Long term drift (1 year) :
±0.10°C / ±0.18°F
Calibration accuracy (test limit) ±0.10°C / ±0.18°F
Temperature coefficient
±0.005°C/°C (0-20°C and 26-40°C) / ±0.009°F/°F