142 2014-07-30 127139 13
THERMAL SPECIFICATIONS –
DRY – BLOCK
RTC-250 A/B
Well specifications Loaded with 2 x 3mm sensors:
40 mm / 1.57 inch axial homogeneity:
±0.030°C/0.054°F@ 28°C/82.4°F to 150°C/302°F
40 mm / 1.57 inch axial homogeneity:
±0.050°C/0.09°F @ 150°C/302°F to 250°C/482°F
50 mm / 1.97 inch axial homogeneity:
±0.050C/0.09°F @ 28°C/82.4°F to 150°C/302°F
50 mm / 1.97 inch axial homogeneity:
±0.100°C/0.18°F @ 150°C/302°F to 250°C/482°F
60 mm / 2.36 inch axial homogeneity:
±0.060°C/0.11°F@ 28°C/82.4°F to 150°C/302°F
60 mm / 2.36 inch axial homogeneity:
±0.120°C/0.22°F @ 150°C/302°F to 250°C/482°F
70 mm / 2.76 inch axial homogeneity:
±0.100°C/0.18°F@ 28°C/82.4°F to 150°C/302°F
70 mm / 2.76 inch axial homogeneity:
±0.200°C/0.36°F @ 150°C/302°F to 250°C/482°F
80 mm / 3.15 inch axial homogeneity:
±0.120C/0.22°F @ 28°C/82.4°F to 150°C/302°F
80 mm / 3.15 inch axial homogeneity:
±0.250°C/0.45°F @ 150°C/302°F to 250°C/482°F
Loaded with 3 x 3mm sensors + 1 x 4mm sensor
and 1 x 6 mm sensor
DLC active:
40 mm / 1.57 inch axial homogeneity:
±0.040°C/0.072°F@ 28°C/82.4°F to 150°C/302°F
40 mm / 1.57 inch axial homogeneity:
±0.060°C/0.11°F @ 150°C/302°F to 250°C/482°F
50 mm / 1.97 inch axial homogeneity:
±0.050°C/0.09°F @ 28°C/82.4°F to 150°C/302°F
50 mm / 1.97 inch axial homogeneity:
±0.070°C/0.13°F @ 150°C/302°F to 250°C/482°F
60 mm / 2.36 inch axial homogeneity:
±0.060°C/0.11°F @ 28°C/82.4°F to 150°C/302°F
60 mm / 2.36 inch axial homogeneity:
±0.100°C/0.18°F @ 150°C/302°F to 250°C/482°F