150 2014-07-30 127139 13
THERMAL SPECIFICATIONS RTC-700 A/B/C
Difference between borings :
±0.01°C/0.02°F @ 33°C/91.4°F to 100°C/212°F
±0.025°C/0.045°F @ 100°C/212°F to
420°C/788°F
±0.040°C/0.072°F @ 420°C/788°F to
600°C/1112°F
±0.050°C/0.09°F @ 600°C/1112°F to
660°C/1220°F
±0.060°C/0.108°F @ 660°C/1220°F to
700°C/1292°F
Influence from load :
0.01°C/0.02°F @ 33°C/91.4°F to 100°C/212°F
0.05°C/0.09°F @ 100°C/212°F to 420°C/788°F
0.15°C/0.27°F @ 420°C/788°F to 700°C/1292°F
Influence from load with ext. reference :
0.01°C/0.02°F @ 33°C/91.4°F to 100°C/212°F
0.02°C/0.04°F @ 100°C/212°F to 420°C/788°F
0.03°C/0.05°F @ 420°C/788°F to 700°C/1292°F
Difference between inserts :
0.05°C/0.09°F @ 420°C/788°F
0.10°C/0.18°F @ 700°C/1292°F
Long term drift (1 year): ±0.11°C/±0.20°F
Calibration accuracy (test limit) ±0.10°C (±0.18°F)
Temperature coefficient ±0.005°C/°C (0-20°C and 26-40°C) / ±0.009°F/°F
(32-68°F and 79-104°F)
Stability ±0.008°C/0.014°F@33°C/91.4°F to 125°C/257°F
±0.015°C/0.027°F@125°C/257°F to 425°C/797°F
±0.020°C/0.36°F@425°C/797°F to 700°C/1292°F
Accuracy ±0.30°C/0.54°F @ 33°C/91.4°F to 660°C/1220°F
±1.69°C/3.04°F@660°C/1220°F to 700°C/1292°F
Heating time incl. insert 33°C/91.4°F to 700°C/1292°F : 45 min.