connected to a mains outlet with a protective earth
connection. Ensure the ground connection of the
calibrator is properly connected to the protective earth
before switching on the calibrator. Always use a mains
power cable with a mains plug that connects to the
protective earth.
To ensure the connection to protective earth any
extension cord used must also have a protective earth
conductor.
Only use a mains power cord with a current rating as
specified by the calibrator and which is approved for the
voltage and plug configuration in your area.
Before switching on the calibrator make sure that it is
set to the voltage of the mains electricity supply.
Always position the calibrator to enable easy and quick
disconnection of the power source (mains inlet socket).
Liquid baths should only be operated by trained
personal.
AMETEK Denmark A/S does not take any
responsibility, if the well is filled with other fluids than
those recommended.
Heat transfer fluids must only be used in calibrators
with a liquid bath. If these fluids are heated above
specified temperature they will create noxious or toxic
fumes. Proper ventilation must be used.
Product information on the fluid must be carefully
investigated before use.
The calibrator must be kept free within an area of 20
cm on all sides and 1 metre above the calibrator due to
fire harzard.
Note…
The instrument must not be exposed to draughts.