127139 13 2014-07-30 69
Results during a switch test
Access the Switch Test Result List by selecting “Result
1
from the Switch Test menu.
This shows the results that are currently available. These
results change as the test progresses.
Press to return to the switch test.
Finished switch test results
At the end of a switch test the results are displayed. These show the
temperature when the thermostat has closed and the temperature
when it has opened – whichever comes first. The difference between
these 2 temperatures is calculated as the hysteresis.
Select “Save
1
” to save the results storing them in the
calibrator’s memory.