16308-101 Rev. A • 18
Instructions for Use (continued)
Understanding Results
High Confidence Final IOP Result
Once enough measurements are taken for a high confidence Final
IOP Result, the Final IOP Result will be briefly displayed in a large
green ring. After, the measurement screen will reappear and the
corresponding OD or OS circle will have a green ring around it.
Low Confidence Final IOP Result
If the Manual Measure Button was pressed during ActiView Auto
Measure Mode, the Final IOP Result will be briefly displayed in a
yellow ring. After, the measurement screen will reappear and the
corresponding OD or OS circle will have a yellow ring around it.
If the other eye is selected before sufficient measurements
were taken to obtain a High Confidence Final IOP Result, the
corresponding OD or OS circle will appear in yellow.
Retake Measurement
If there is a high variability between measurements (high standard
deviation) the Retake Measurement arrow will appear with a large
orange ring around it. Retake the measurement to obtain a reliable
Final IOP Result.
NOTE: The Retake Measurement arrow will not apear when
Measurement Mode is set to 1 (Quick).