Slip blocks
The gauge can be calibrated by taking measurements over a set of slip 
blocks. The slip blocks artefact is defined in the Instrument Configuration 
(Gauge > Artefacts). Calibration procedure steps are below:
1.  Navigate to the ‘Gauge’ section of the instrument ribbon and click on 
the ‘Calibration’ icon. 
2.  The calibration dialog box will then be displayed. Select a slip block 
calibration artefact form the drop-down list. Complete the relevant 
options (Explained on Page 30) and click ‘Next’.
1
2
5
4
3.  Follow the on-screen instructions. Click ‘Next’ to continue through 
the calibration procedure.