4
English
Start up procedure
Note
Please follow the optimised start up procedure to ensure quick
and easy start up of Talyrond
®
500 PRO and Metrology 4.0:
1. Switch on the external air supply.
If instrument still indicates ‘Low air pressure’ when the software
starts, check that the air is ‘On’ on the internal air panel.
2. Switch on either.
Distribution unit (Electrical cabinet only)
ECU (Electronic control unit only)
3. Switch on the PC and monitors.
4. Switch on Active AV
Systems with Active AV only (e.g. Talyrond
®
595H PRO)
Introduction
The Talyrond
®
500 PRO is a range of fully automated roundness/
cylindricity instruments that are unsurpassed in accuracy and reliability.
The instruments has been designed for stability and accuracy and consists
of a frictionless air bearing spindle mounted within a cast iron base and a
precision machined column.
The world-leading gauge delivers 4 mm of range, twice the range of most
roundness instruments. This enables the unique capability of roundness,
contour, and roughness measurement from a single gauge. This universal
gauge has a resolution down to 0.3 nm, when coupled with the system’s
low noise platform, the Talyrond
®
500 PRO becomes a fully automated
surface texture instrument.
All installation and servicing should be carried out by a Taylor Hobson
accredited service engineer.
The use of these instruments for any other purpose than a metrology
measuring instrument is not warranted by Taylor Hobson. Be aware of all
relevant safety regulations.