VersaSTAT LC Low Current Interface (LCI) Hardware Manual P/N 224186_0B
6
Figure 6. Calibration Required Message
This message will appear after adding the LCI option to a system, as well as after removing the LCI
option.
The calibration procedure is located in the Tools>Options… menu within VersaStudio. Once the LCI is
properly connected and configured for the experimental conditions, select the “Calibrate Now” button.
Figure 7. Calibration Procedure
NOTE It is very important that the calibration be performed in the exact same
manner as the experimental settings. The lowest current ranges and current
measurements are affected by temperature fluctuations, vibrations, and
environmental sources of electromagnetic fields. Always calibrate the
system in the same experimental setup as your experiment requires.
Also, it is very important that the LCI and potentiostat be allowed to warm
up and reach temperature equilibrium. After power-on, wait a minimum of
20 minutes BEFORE performing the calibration.