Figure 13. Calibration Complete Message 
 
The calibration will not take effect until the system is re-booted by cycling the power on the potentiostat. 
 
 
 
 
 
2.3. LCI Checkout Procedure 
As shown in Section 1.2.2, the VersaSTAT LC is equipped with a built-in 10Mohm resistor that can be 
used as a “dummy cell” to perform a limited test of the system for proper operation. 
 
To test the LCI with a VersaSTAT using dc voltage scan: 
1. 
Connect the cell leads to the built-in 10Mohm dummy cell (The WE cable connects to the W pin 
jack, the SE cable connects to the S pin jack, the RE cable connects to the R pin jack and the CE 
cable connects to the C pin jack). 
2. 
Select  “New  Experiment”  in  VersaStudio  and  choose  the  Staircase  Linear  Scan  Voltammetry 
action.  Set  the  Initial  Potential  =  -100mV,  Final  Potential  =  +100mV,  Step  Height=1mV,  Step 
Time=1s,  Acquisition  Mode=Average,  Electrometer=Single  Ended,  E  and  I  Filters  =  10Hz  (as 
shown below).  Once setup, select “Run” to acquire the data. 
3. 
Once the data collection has finished, select all of the data points (Data>Select All) and perform a 
Line Fit by selecting the Line Fit button at the top of the graph. 
4. 
The vertex currents should be +/-10nA, and the slope should be 1.0E-7 (the inverse of which is 
10Mohm). 
 
The PARSTAT4000 + VersaSTAT LC combination accesses additional lower current ranges than when 
used with a VersaSTAT. These lower ranges have longer settling times, so the Checkout Procedure is 
slightly different. If you use the VersaSTAT + VersaSTAT LC Check Out procedure on this combination, 
data around 0 Amps will show a large measurement artifact (spike). 
 
To test the LCI with a PARSTAT4000 using dc voltage scan: 
1. 
Connect the cell leads to the built-in 10Mohm dummy cell. (The WE cable connects to the W pin 
jack, the SE cable connects to the S pin jack, the RE cable connects to the R pin jack and the CE 
cable connects to the C pin jack.) 
2. 
Select  “New  Experiment”  in  VersaStudio  and  choose  the  Staircase  Linear  Scan  Voltammetry 
action. Set the Initial Potential = -100mV, Final Potential = +100mV, Step Height = 3.5mV, Step 
Time = 7s, Acquisition Mode = Average, Electrometer = Single Ended, E and I Filters = 10Hz (as 
shown below).  Once setup, select “Run” to acquire the data.