7 SPECIFICATIONS
7.1 Electrochemical O
2
Mechanical and Environment
15% to 90% r.h. non-condensing
Response Time (t90)
via diffusion
Stabilization time
(for calibration
adjustments)
±0.3% v/v over the range 10% to 25% v/v
±0.5% v/v over the range 0% to 9.9% v/v
1
The output measurement resolution is consistent across all gas types and does not infer the accuracy of
the gas type
Table 9: Typical Oxygen Cross-Sensitivity
Instrument variance from
Standard Test Gas
10ppm Hydrogen Sulphide in Air
10% Carbon Dioxide in Air
5% Nitrogen Dioxide in Air