Mechanical and Environment
15% to 90% r.h. non-condensing
Response Time (t90)
via diffusion
<5 mins (60 seconds typical)
Stabilization time
(for calibration
adjustments)
±3 ppm over the range 0 ppm to 29 ppm
±6 ppm over the range 30 ppm to 100 ppm
1
The output measurement resolution is consistent across all gas types and does not infer the accuracy of
the gas type
Table 10: Typical Carbon Monoxide Cross-Sensitivity
Instrument variance from
Standard Test Gas
50ppm CO + 5% Methane in Air
50ppm CO + 0.01% Hydrogen in Air
50ppm CO + 10% Oxygen in Nitrogen
50ppm CO + 10 ppm Hydrogen Sulphide in Air
50ppm CO + 5% Carbon Dioxide in Air
50ppm CO + 5% Nitric Oxide in Air
50ppm CO + 5% Nitrogen Dioxide in Air