EasyManua.ls Logo

Ampcontrol Gasguard 2 GG2 DETECTOR - NDIR CH4 and CO2 Specifications

Ampcontrol Gasguard 2 GG2 DETECTOR
39 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ampcontrol Pty Ltd ABN 28 000 915 542
GG2 DETECTOR USER MANUAL
GG2B011
Version 3 June/19
Uncontrolled Copy - Refer to Ampcontrol Website for Latest Version
APPROVED FOR EXTERNAL DISTRIBUTION
PROPERTY OF AMPCONTROL PTY LTD
NOT TO BE
REPRODUCED IN PART
7.3 NDIR CH
4
Electrical
Supply Voltage
10 - 16.5VDC
Max Power
Consumption
0.1W
Mechanical and Environment
Dimensions (H x W x D)
86.5mm x 85mm x 93mm
Weight
1kg
IP Rating
IP54
Humidity Range
15% to 90% r.h. non-condensing
Temperature Range
-20˚C to + 40˚C
Pressure Range
±20kPa
Response Time (t90)
via diffusion
<30s
Warm-up Time
<5 mins (60 seconds typical depending on ambient temperature)
Stabilization time
(for calibration
adjustments)
3 mins
Sensor Unit
Detection Method
Non-dispersive infrared
Calibrated Range
0-5% v/v
Accuracy
±0.1% v/v or ±5% of indication
1
Drift
±0.1% v/v or ±5% of indication
1
over 3 months
Repeatability
±0.1% v/v or ±5% of indication
1
Sensing Element Life
>5 years
Resolution
0.01% v/v
2
Certification
IECEx
IECEx TRA 16.0020X
DPI
MDR 0001030 GD
1
Whichever is greater
2
The output measurement resolution is consistent across all gas types and does not infer the accuracy of
the gas type
Table 11: Typical NDIR Methane Cross-Sensitivity
Test Gas Mixture
Instrument variance from
Standard Test Gas
2% Methane + 13% Oxygen in Nitrogen
±0.2% Methane
2% Methane + 5% Carbon Dioxide in Air
±0.2% Methane
2% Methane + 0.075% Ethane in Air
±0.2% Methane
2% Methane + 0.020% Ethane in Air
±0.2% Methane