38 US/50 User's Guide
Part 5
17. Slowly increase the Ultrasound Intensity control to the desired output level
as indicated on the ultrasound meter dial. If desired, adjust the Ultrasound
Interrupted Output slide control now. When the Ultrasound Interrupted
Output slide control is adjusted, the Interrupt On indicator light (located
above the upper right corner of the ultrasound meter) will illuminate and the
Ultrasound Meter indicator will rise and fall indicating interrupted cycles
provided the transducer is not in the ultrasound cradle. Stimulator output is
unaffected.
18. Slowly increase the Low Volt AC Intensity control for Stimulator Output to
the desired output level.
Note: If the Stimulator Reset Intensity indicator light is illuminated, no output
will be obtained until the Intensity control for Stimulator Output is returned
to the 0/Reset position enabling the audible "click".
19. Select an alternate Output Mode, Pulsation, or Surge, and adjust the Pulse/
Surge Rate control if so desired. The Output Rate indicator light will reflect
the output rate selected.
20. When treatment is completed, the ac power will shut off and a bell will sound.
To continue or extend a patient's treatment time, reset the Power/Timer knob
within ten (10) seconds of shut off. To initiate early shut off, turn the Power/
Timer knob counterclockwise to the Off position. The ac power will shut off
and a bell will sound. Reduce the Stimulator Intensity control for Stimulator
Output to the 0/Reset position enabling the audible "click". Reduce the
Ultrasound Intensity control to "0" position and thoroughly clean the
transducer faceplate before placing the transducer in the transducer cradle.
Thoroughly clean the pad with warm water and dispose of the used cloth
cover.