O2NE+ & Safe-Ox+
User Manual Int. Approved
Document ref: P0120-820-14 August 2017 Page 18 of 23
Commercial in Confidence
7 Specifications
Better than ± 0.75% O2 over 5.0 to 25.0% O2
Better than ± 1.00% O2 over 0.1 to 5.0% O2
0 to +40
°
C (+32 to +104
°
F)
0.2% of reading/
°
C or 0.1115% of reading/°F
Atmospheric pressure range
10 seconds to normal operation
Prior to calibration allow 2 hours to achieve full accuracy.
Hard wired – IP65
Quick connect – IP43
4 digit Liquid Crystal Display
2 x Alarm Visual Indicators
1 x System Fault Indicator
1 x Status Indicator
Common Audible Alarm
One or Two Optional Alarm Relays with changeover contacts assigned
to Alarm 1, Alarm 2 or System Fault.
Contact Rating 240V AC or 30V DC at up to 2A.
Contacts are non-latching Fail-Safe.
2 wire, 4 to 20mA (Max load 150 Ω)
a) 210 to 250VAC supply
b) 110 to 120VAC supply
c) 9-24VDC supply
* Typical performance, with fixed temperature, humidity and pressure