O2NE+ & Safe-Ox+
User Manual Int. Approved
Document ref: P0120-820-14 August 2017 Page 3 of 23
Commercial in Confidence
4 to 20mA (or 0-1V) Output:-
External connected equipment must meet the requirements for reinforced insulation or must be
a Safety Extra Low Voltage (SELV) circuit as defined in IEC60950-1.
The above statement refers to the risk of externally connected equipment introducing
hazardous voltages into the A50 (O2NE+, Safe-Ox+ or CO Clear). It is the
responsibility of users of this product to ensure that their equipment/accessory is safe before
connecting it.
NOTE - If the equipment is used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.
NOTE - The O2NE+ and Safe-Ox+ range of instruments are precision instruments and
therefore are susceptible to shock. Care should be taken when installing your O2NE+ or
Safe-Ox+ to avoid instrument damage.