EasyManua.ls Logo

Andor Technology Zyla sCMOS 5.5 - Rolling Shutter Internal Triggering (Overlap Mode)

Default Icon
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
version 2.0 rev 18 Feb 2021
33
2.6.2.4 Rolling Shutter Internal Triggering (Overlap Mode)
Internal Triggering in Overlap Mode allows the user to perform an exposure and acquire images from the
sensor simultaneously. This is achieved by starting a new exposure for a new frame while the current frame’s
exposure is being read out from the sensor. The cycle time is the same as the exposure time entered by the
user in this scenario.
Short Exposures
When the required exposure time is less than the time it takes to read out a frame (Short Exposures), the cycle
timeisalwaysdenedbythetimetakentoreadoutaframe.
Figure 17: Rolling Shutter Internal Triggering (Short Exposures)
Table 13: Rolling Shutter Internal Triggering (Overlap Mode- Short Exposures) Timing Parameters
Parameter Minimum Maximum
Exposure 1 Row 1 Frame - 1 Row
Cycle Time (1/Frame Rate) 1 Frame 1 Frame + Exposure + 1 Row
FIRE any low period 1 Row -